Return to Axis ORbits

More Accurate Execution? Preoperative Planning vs. Planning + GPSTM Computer Navigation

GPS Navigation
Share this article

Study Question

Does GPS computer navigation lead to more accurate glenoid positioning than CT-based preoperative planning alone, regardless of a surgeon’s experience level?

The Story

3D CT-based planning software is the gold standard for preoperative planning for shoulder arthroplasty. However, planning is not the same as execution. Schoch and colleagues compared 50 cases to see how well surgeons of different experience levels (fellows vs. attending surgeons) re-created their pre-op plans when blinded to the navigation screen. They discovered that even for experienced, high-volume surgeons, the gap between the plan and reality was surprisingly wide.

Key Findings

  • An execution gap: Surgeons were unable to reliably execute their preoperative plans without computer navigation, with malposition (>4 mm of displacement or >10° error in version or inclination) occurring in 38% of cases by attending surgeons and 58% of cases for fellows.
  • The power of experience: While fellow trainees had significantly higher malposition rates (p=0.047), even fellowship-trained attendings struggled to re-create intended baseplate tilt and version. Without computer navigation, baseplates were implanted in >10° superior/inferior inclination deviation in plan by 14% of attendings’ and 35% of fellows’ cases (p=0.057).  Similarly, baseplates were implanted in >10° anteversion/retroversion deviation in plan by 18% of attendings’ and 31% of fellows’ cases (p=0.21).
  • The case for navigation: The mean offset from the planned starting point on the glenoid was 3.2±2.0mm without computer navigation. These high execution errors support the need for computer navigation to ensure a surgeon’s 3D surgical plan is achieved in the OR.
Errors by Type & Experience

Deviation to Plan without GPS (% of cases off >10°)

Superior/Inferior Inclination
Fellows35%
35%
Attending Surgeons14%
14%
Anteversion/Retroversion
Fellows31%
31%
Attending Surgeons18%
18%

Clinical Context

3D planning is only half the battle. Without computer navigation, there is a significant risk that the carefully crafted pre-op plan won’t be realized, which could compromise the long-term success of the replacement.

Limitations

These findings are limited because the study design only investigated 50 patients and utilized a small group of surgeons at a single academic center.

Reference for this study: Schoch BS, et al. Computer navigation leads to more accurate glenoid targeting during total shoulder arthroplasty compared with 3-dimensional preoperative planning alone. J Shoulder Elbow Surg. 2020;29(11):2257-2263. doi: 10.1016/j.jse.2020.03.014.

GPS, Shoulder Planning App and Predict+ are developed by Blue Ortho SAS, an Advita Ortho subsidiary, and distributed by Advita Ortho, LLC.

3.2 ± 2.0mm

Mean Offset
from planned glenoid starting point without computer navigation

The Advita multicenter Axis research group efforts make these generalizable findings possible.

Share this article

Will add hubspot form for signup here. This is placeholder

Read More Articles:

Subscribe to get the latest OR-bits – our clinical updates, published data, and concise research insights from the Axis program.